Yokogawa Electric SC92-21-J Portable Conductivity Meter

Yokogawa Electric SC92-21-J Portable Conductivity Meter

Manufaturer: Yokogawa Electric
Model: SC92-21-J
Product code
111-67402
Price 1,373.26 USD
Add to cart Buy now

*The above price is Standard Price, Kindly click on Request a quote to unlock the Best Price and Delivery Date. Our expert consulting team is always ready to assist you.

*Order Process: 1. Request a quote ⇒ 2. Quote via Email ⇒ 3. Proforma Invoice ⇒ 4. Payment ⇒ 5. Delivery and Invoice.

*You can't find the product you need, please send us the picture/specifications/model/product code and quantity you need to email: [email protected]. We will find it for you.

*Review us here: Review Now 

Description

  • The stability judgment function supports accurate measurements.
  • The detector can be attached to the main unit.
  • It is convenient to carry around.
  • We have pursued ease of use with features such as a one-touch calibration button.
  •  
  • Measurement target: Conductivity or resistivity of aqueous solution
  • Measurement range: Conductivity (min. to max. range) / (when connected to pure water detector) 0-2.00 µS/cm to 0-200.0 µS/cm (when connected to general-purpose detector) 0-20.00 µS/cm to 0-200.0 mS/cm (when connected to highly corrosion-resistant detector) 0-20.00 µS/cm to 0-200.0 mS/cm (when connected to high concentration detector) 0-200.0 µS/cm to 0-2.000 S/cm, Resistivity / 0.001 MΩ・cm to 40.0 MΩ・cm (for pure water detector) Temperature: 0 to 80 °C (detector combination) * 0 to 50 °C when the entire detector is submerged
  • Number of digits displayed: Conductivity/Maximum 4 digits (automatic range switching), Resistivity/Maximum 4 digits
  • Temperature display range: -10.0 to 120.0°C
  • Repeatability (main unit and detector combination): ±2% of full scale conductivity (however, for the 0-200 mS/cm range of general-purpose detectors, ±5% of full scale)
  • Temperature accuracy: Simulated input of main unit only: ±0.5°C ±1 digit; detector combination: ±1.0°C (0 to 70°C), ±1.3°C (over 70°C)
  • Display: Digital LCD display + backlight
  • Temperature compensation: 0.00 to 10.00%/℃ (reference temperature conversion can be set between 15 to 30℃)
  • Calibration: Conductivity/Resistivity/Auto (up to 2 points)/Manual (up to 5 points), Temperature/Manual (1 point)
  • Functions: Simultaneous display of conductivity or resistivity and liquid temperature, electrode quality diagnosis function, stability judgment function, data memory (500 items)
  • Structure: IP67
  • Compliant standards: CE, RCM, KC
  • Compatible detector: SC92SN
  • Wetted parts materials: Pure water detector/SUS316 (electrode element), polypropylene resin (insulating detector body), fluororubber (O-ring), PVC (cable), silicone rubber (detector seal), hard polyethylene (grip), nylon 6 (cord bushing), epoxy resin (detector internal filler); General-use detector/Titanium (detector electrode), polypropylene resin (detector body, cover), fluororubber (O-ring), PVC (cable), polyphenylene sulfite resin (constriction), silicone rubber (detector seal), hard polyethylene (grip), nylon 6 (cord bushing), epoxy resin (detector internal filler); High corrosion resistance/high concentration detector/Glass (detector body), platinum black (electrode element), PVC (cable), silicone rubber (detector seal), hard polyethylene (grip), nylon 6 (cord bushing), epoxy resin (detector internal filler)
  • Ambient temperature: 0-45°C
  • External dimensions: Approx. H160 x W80 x D40 mm (excluding protrusions)
  • Weight: Approx. 230 g (excluding batteries and detector)
  • Power supply: Alkaline batteries (2 AA batteries), variable auto power off function (1 to 30 minutes) Battery life: Approximately 500 hours
  • Frequently Asked Questions for the Yokogawa Electric SC92-21-J Portable Conductivity Meter:

    1. What is the SC92-21-J and what are its typical applications?

    The SC92-21-J is a handheld portable conductivity meter designed by Yokogawa for measuring the conductivity of liquids in the field or laboratory. It provides reliable measurements from low-conductivity water to high-salinity solutions, making it suitable for water quality monitoring, environmental testing, and industrial process control.

    2. What measurement ranges and units does it support?

    The SC92-21-J covers a broad conductivity range of 0 to 2.00 S/cm and low-conductivity measurements down to 0 to 2.00 µS/cm. It also offers resistivity readings up to 200 MΩ·cm. Users can switch display units between µS/cm, mS/cm, S/cm, and Ω·cm directly on the device.

    3. How accurate is the SC92-21-J and what resolution does it offer?

    Typical accuracy for the SC92 series is ±1% of the reading, depending on the selected measurement range. Resolution can be as fine as 0.01 µS/cm in the lowest range and 0.001 mS/cm in higher ranges, ensuring precise readings across various sample types.

    4. What safety precautions should I follow when using the SC92-21-J?

    Do not use the device in environments with the risk of electrical shock and avoid touching the electrode immediately after immersion in very hot solutions to prevent burns. Handle the glass sensor carefully to avoid breakage and observe electrostatic discharge precautions when inserting batteries or connecting sensors.

    5. How do I set up and operate the meter for basic measurements?

  • Install two AA batteries and power on the device.

  • Select the desired measurement unit.

  • Rinse the sensor with distilled water and immerse it fully into the sample.

  • Gently stir or tap the sensor to release any trapped air bubbles.

  • Wait for the stability indicator to appear on the screen, then record the measurement.

  • Rinse the sensor again and store it according to the manual instructions.

  • 6. How often should I calibrate the SC92-21-J and how is it done?

    Calibration should be performed regularly using manufacturer-supplied standard solutions. Enter calibration mode, immerse the sensor in the calibration solution, wait for stability, and confirm the calibration when the “OK” or stability symbol appears. Always follow the user manual’s detailed calibration procedure for best results.

    7. What is the expected battery life and how should I handle battery disposal?

    With two AA batteries installed, the meter can operate continuously for approximately 200 hours, depending on backlight usage and measurement frequency. Remove batteries when the device will not be used for extended periods to prevent leakage. Batteries must be disposed of in compliance with local regulations and environmental standards.

    8. How do I maintain and store the SC92-21-J to ensure longevity?

    After each use, clean the sensor with distilled water or the recommended cleaning solution and dry it gently with a soft cloth. Store the meter and sensors in the provided carrying case in a cool, dry place. Remove batteries if the device will be unused for long periods to avoid damage from leakage.

    9. Where can I find the complete user manual and warranty information?

    The full user’s manual for the SC92 series, including the SC92-21-J model, is available online (document IM 12D03E02-02EN) and please contact QTE Technologies (qtetech.com/en). The product comes with a 12-month warranty covering defects in materials and workmanship, excluding consumable parts like electrodes and batteries.

    10. What accessories and sensors are compatible with the SC92-21-J?

    The SC92-21-J accepts standard conductivity sensors from Yokogawa, including general-purpose probes for industrial water and high-sensitivity probes for ultrapure water. Accessories include calibration solution kits, protective caps, and carrying cases designed to fit the meter and probes.

Genuine commitment
Genuine commitment
Flexible payment
Flexible payment
Dedicated and free support
Dedicated and free support
Domestic and international delivery
Domestic and international delivery
Always best price
Always best price

You may like

Sign up for the newsletter - the opportunity to receive promotions